DISTRIBUTOR FLOW METER TOKICO

Pressure Gauge Model 113.53


Description

Design

EN 837-1

Nominal size in mm

40, 80, 100

Accuracy class

2.5

Scale ranges

NS 40: 0 … 1 to 0 … 400 bar

NS 80, 100: 0 … 1 to 0 … 600 bar

or all other equivalent vacuum or combined pressure  and vacuum ranges

Pressure limitation

Steady: 3/4 x full scale value

Fluctuating: 2/3 x full scale value

Short time: Full scale value

Permissible temperature

Ambient: -20 … +60 °C

Medium: +60 °C maximum

Temperature effect

When the temperature of the measuring system

deviates from the reference temperature (+20 °C):

max. ±0.4 %/10 K of the span

Ingress protection

IP 65 per EN 60529 / lEC 60529

 

APPLICATION

  • For measuring points with high dynamic pressure loads or vibrations
  • For gaseous and liquid media that are not highly viscous or crystallising and will not attack copper alloy parts
  • Hydraulics
  • Compressors

SPECIAL FEATURES :

  • Vibration and shock resistant
  • Robust design
  • Stainless steel with liquid filling
  • Scale ranges up to 0…600 bar

Have any Question or Comment?

Leave a Reply

Your email address will not be published. Required fields are marked *

Copy Protected by Chetan's WP-Copyprotect.